Uncertainty budget of temperature measurement by the methods of monochromatic one- and two-color thermometry with emissivity correction
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Видавництво Львівської політехніки
Lviv Politechnic Publishing House
Lviv Politechnic Publishing House
Abstract
The work is directed on accuracy increasing of the thermodynamic temperature measurement of heated objects by
their own thermal radiation, utilizing no more than two monochromatic channels for the spectral density of radiance measurement.
The classification of the methods of one- and two-color thermometry with emissivity correction have been proposed. It has
been determined that these methods are based on generalized measurement equation originated from the spectral density of radiance
approximation by the Wien’s formula. Temperature measurement equations have been reformulated basing on the set of common
input quantities. This set includes conditional one-color (radiance) temperatures, wavelengths and correction parameters, that account
for spectral emissivity of controlled object. The number of input quantities for one-color methods is 4, and for the most part of twocolor
methods – 6.
Taking into consideration the specific of hardware realization of thermometry system (microspectrometer with multielement
CCD-sensor of radiation), uncertainty components of input quantities have been researched and systematized, as well as the formulas
for input quantities uncertainties evaluation basing on these components have been obtained. The uncertainty budget of indirect
temperature measurement by the methods of monochromatic one- and two-color thermometry has been formed on the base of the
input quantities set. The formulae for evaluation of combined standard and expanded, with confidence level 0.95, uncertainties of
temperature measurement result have been obtained. They account for the possible correlation between correction parameters. For
expanded uncertainty evaluation the kurtosis method has been utilized. It has been shown, that under conditions of the same
uncertainties of input quantities, the accuracy of thermometry methods depends on the corresponding sensitivity coefficients only. The
necessity of obtaining analytical expressions for sensitivity coefficients have been argued. This will allow analyzing the complex
influence of input quantities uncertainties on temperature measurement accuracy by investigated thermometry methods, in order to
determine the method, which demonstrates minimal possible uncertainty under specified measurement conditions.
Description
Citation
Petrenko D. Uncertainty budget of temperature measurement by the methods of monochromatic one- and two-color thermometry with emissivity correction / Dmytro Petrenko, Oleksandra Hotra // Measuring Equipment and Metrology. — Lviv : Lviv Politechnic Publishing House, 2025. — Vol 86. — No 2. — P. 18–26.