Emission current stabilizer to electron-impact gas ion sources

dc.contributor.authorSikora, J.
dc.contributor.authorToborek, K.
dc.contributor.authorSzczepaniak, L.
dc.date.accessioned2016-01-21T13:09:30Z
dc.date.available2016-01-21T13:09:30Z
dc.date.issued2004
dc.description.abstractWe describe an electron emission current stabilizer with an error amplifier dc voltage gain depending on the electron emission current. A negative feedback loop of the error amplifier contains a solid-state digital potentiometer, which is controlled by a microprocessor according to the empirically determined (based on the Ziegler - Nichols method) exponential dc voltage gain function. The obtained mean relative standard deviation of the electron emission current, for a tungsten filament operated at the pressure of/?=0.1Pa, is 36 ppm.uk_UA
dc.identifier.citationSikora J. Emission current stabilizer to electron-impact gas ion sources / J. Sikora, K. Toborek, L. Szczepaniak // Вісник Національного університету «Львівська політехніка». – 2004. – № 510 : Елементи теорії та прилади твердотілої електроніки. – С. 112–114. – Bibliography: 4 titles.uk_UA
dc.identifier.urihttps://ena.lpnu.ua/handle/ntb/31098
dc.language.isoenuk_UA
dc.publisherВидавництво Національного університету "Львівська політехніка"uk_UA
dc.titleEmission current stabilizer to electron-impact gas ion sourcesuk_UA
dc.typeArticleuk_UA

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
23_112-114.pdf
Size:
166.38 KB
Format:
Adobe Portable Document Format