Investigation of ultrasonic flowmeter error in conditions of distortion of flow structure
Loading...
Date
2015
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Lviv Polytechnic Publishing House
Abstract
Ultrasonic flowmeter error caused by distortion of flow structure has been investigated in this paper.
Minimum number of acoustic chordal paths of multi-channel ultrasonic flowmeter was determined that ensures the error caused by distortion of flow structure less than the specified limits ±0.3 %. The authors used Salami velocity function as a model of distortion of flow structure. Acoustic path
configurations of ultrasonic flowmeter were determined using classical numerical methods of integration of Gauss-Jacobi. The authors improved the flow rate equation of ultrasonic
flowmeter by introduction of calibration coefficient that was determined using symmetric component of Salami function.
Description
Keywords
ultrasonic flowmeter, flow structure, local fitting, Salami function, numerical method of integration, acoustic path, calibration coefficient
Citation
Roman V. Investigation of ultrasonic flowmeter error in conditions of distortion of flow structure / Vitalii Roman, Fedir Matiko // Litteris et Artibus : proceedings of the 5th International youth science forum, November 26–28, 2015, Lviv, Ukraine / Lviv Polytechnic National University. – Lviv : Lviv Polytechnic Publishing House, 2015. – P. 190–191. – Bibliography: 5 titles.