Study of the effective radii of ultrasonic transducers
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Видавництво Національного університету “Львівська політехніка”
Lviv Politechnic Publishing House
Lviv Politechnic Publishing House
Abstract
The study on the effective radii of ultrasonic transducers for hydrophone calibration at the National State Primary Standard of the Unit of Ultrasonic Pressure in Water (NDETU AUV-02-2018) is presented and the appropriate method is described. The impact of the effective radius on measurement distance and diffraction loss coefficients is evaluated. The uncertainty calculation of the effective radius measurement of ultrasonic transducers is provided, and its influence on the accuracy of hydrophone calibration is assessed. Considerable attention is given to estimating the measurement distance between the hydrophone and the ultrasonic transducer, which significantly affects the accuracy of hydrophone calibration.
Description
Citation
Shpak O. Study of the effective radii of ultrasonic transducers / Oleksandr Shpak, Dariia Duviria // Measuring Equipment and Metrology : scientific journal. — Lviv : Lviv Politechnic Publishing House, 2024. — Vol 85. — No 3. — P. 10–16.